(i) at pH between 5.0 and 5.5?
(1)
(ii) at pH 3.5?
(1)
(b) Figure (b) is also called a window diagram and is used to find the optimum
separation by plotting a for each pair of solutes. Using this figure, explain what
the optimum mobile phase pH would be to obtain the best chromatographic
separation of these 4 components.
(4)
3.3 The following figure shows the effect of flow rate on the column's plate height.
60
--E
:::1. 40
.c
0)
·a3
a-:.c
(I}
rd
20
0
5 ~~Im-
3µm
2
4
6
8
Flow rate (ml/min)
(a) In the Van Deemter Equation, H = A + B/µ + Cµ, which parameter is affected by
changing the flow rate of a chromatographic system and what is that parameter
called?
(2)
(b) Explain the change (i.e. decreasing or increasing trend) observed in the size
of the plate height when increasing the flow rate.
(2)
(c) Which optimum flow rate would you use to obtain the best efficiency with the
three particle sizes in the figure above and why?
(2)
(d) Assuming you have two columns with particle sizes 3 µm and 10 µm. Which one
would you choose to carry out your analysis (all other characteristics are similar
for both columns) and why?
(3)
3.4 A 1.5 m column has 45000 plates per meter. Analyte A has a retention time of 16.3
minutes, analyte B has a retention time of 17.1 minutes and the dead (or void) time
is 5.14 minutes.
(a) What is the resolution for A and Bon this column?
(4)
(b) What is the selectivity of the column?
(2)
5